Microradiography is an imaging technique using X-rays in the study of internal structures of objects. This rapid and convenient imaging tool is based on differential X-ray attenuation by various tissues and structures within the biological sample. The non-absorbed radiation is detected with a suitable detector and creates a radiographic image. In order to detect the differential properties of X-rays passing through structures sample with different compositions, an adequate high-quality imaging detector is needed. We describe the recently developed radiographic apparatus, equipped with Timepix semiconductor pixel detector. The detector is used as an imager that counts individual photons of ionizing radiation, emitted by an X-ray tube FeinFocus with tungsten, copper or molybdenum anode. Thanks to the wide dynamic range, time over threshold mode — counter is used as Wilkinson type ADC allowing direct energy measurement in each pixel of Timepix detector and its high spatial resolution better than 1μm, the setup is particularly suitable for radiographic imaging of small biological samples. We are able to visualize some internal biological processes and also to resolve the details of insects (morphology) using different anodes. These anodes generate different energy spectra. These spectra depend on the anode material. The resulting radiographic images varies according to the selected anode. Tiny live insects are an ideal object for our studies.
Granty

Projekty


Ocenění
Award for third best poster

Příklad citace článku:
J. Dammer, F. Weyda, J. Benes, V. Sopko, J. Jakůbek, V. Vondracek, "Microradiography of biological samples with Timepix", JINST 6 C11005 doi:10.1088/1748-0221/6/11/C11005 (2011)

Hledat
Události

IEEE NSS/MIC/RTSD
Seoul
27 Oct - 2 Nov 2013

15thIWORID
Paris
23-27 June 2013

NSS MIC IEEE Conference
Anaheim, USA
29 Oct - 3 Nov 2012

Int. Workshop on Radiation Imaging Detectors
1-5 July 2012

Program oslav 10. výročí založení ÚTEF
4-6. 6. 2012

Pozice pro mladé výzkumné pracovníky

Příležitostní razítko k 10. výročí ÚTEF
Na Hrádku 2, Praha 2
30.4, 2-4.5 2012

Collaboration Meeting NEMO/SuperNEMO
Prague, Czech Republic
27.2.-1.3. 2012

Vzpomínkové setkání k připomenutí životní cesty RNDr. Františka Lehara, DrSc.
Refektář Benediktinského opatství v Emauzích, Praha 2
8. února 2012

Zemřel František Lehar
Lyon, Francie
23.11.2011

Vše...
10th Anniversary of IEAP