IEAP - Institute of Experimental and Applied Physics CTU - Czech Technical University in Prague
Česky English
IEAP - Institute of Experimental and Applied Physics CTU - Czech Technical University in Prague
CTU - Czech Technical University in Prague
Partial tasks  > X-ray flourescent imaging
X-ray flourescent imaging


X-ray fluorescence (XRF) is known method for the material elemental composition analysis. Usually, high quality spectroscopic detectors are used for precise elemental analysis, however it is very difficult to obtain information about spatial distribution of elements. To get an image of a spatial distribution of elements contained in the specimen we use the Medipix or TimePix pixel detectors.
In our setup, the tested sample is iluminated by an X-ray tube. The Timepix pixel detector with pinhole aperture is used. Energy resolution of the Timepix device is approx. 3 keV, which is not sufficient for full elemental analysis. If global elemental composition of the specimen is known (or it can be estimated) and if calibration of response of each Medipix pixel to fluorescence spectrum of each of these elements is done, then it is possible to retrieve the distribution of elements in the sample using spectrum decomposition technique.
Medipix2 detects X-rays particles with minimal energy approx. 3keV. That is why it is possible get an image of spatial distribution of elements with atomic number greater than 19 (potasium). It is possible to resolve elements with difference in the atomic number of 1.



Responsible person

Co-workers

Grants:
Number Name Agency
1P04LA211 Collaboration Czech Republic with CERN MŠMT
MSM 6840770040 Research Program 40: Usage of radionuclides and ionizing radiation. MŠMT


Articles in Impacted Journals
(6)
Table format
Year

  • J. Žemlička, J. Jakůbek, F. Krejčí, D. Hradil, J. Hradilová, H. Mislerová, "Mobile system for in-situ imaging of cultural objects", Journal of Instrumentation 7 C01108 (2012)

  • J. Žemlička, J. Jakůbek, M. Kroupa, D. Hradil, J. Hradilová, H. Mislerová, "Analysis of painted arts by energy sensitive radiographic techniques with the Pixel Detector Timepix", Journal of Instrumentation 6 C01066, doi: 10.1088/1748-0221/6/01/C01066 (2011)

  • J. Uher, G. Harvey, J. Jakůbek, "X-Ray Fluorescence Imaging With the Medipix2 Single-Photon Counting Detector", IEEE Transactions on Nuclear Science, Vol. PP, Issue: 99, doi: 10.1109/TNS.2011.2173352

  • J. Žemlička, J. Jakůbek, M. Kroupa, V. Tichy, "Energy and Position Sensitive Pixel Detector Timepix for X-Ray Fluorescence Imaging", NIM A, Vol. 607, Issue 1, p. 202-204 (2009)

  • J. Jakůbek, "Semiconductor Pixel Detectors and their Applications in Life Sciences", 2009 JINST 4 P03013 doi:10.1088/1748-0221/4/03/P03013

  • V. Tichy, T. Holý, J. Jakůbek, V. Linhart, S. Pospíšil, Z. Vykydal, "X-Ray Fluorescence Imaging with Pixel Detectors", Nucl. Instr. and Meth. A, Volume: 591, Issue:1, Pages: 67-70, doi: 10.1016/j.nima.2008.03.122
Search
10th Anniversary of IEAP