Publication  > 'Three-dimensional dosimetry imaging of I-125 plaque for eye cancer treatment'
Three-dimensional dosimetry imaging of I-125 plaque for eye cancer treatment

Author

Year
2011

Scientific journal
Nucl. Instr. Methods A, Vol. 633, Suppl. 1, p. S276-S278, doi:10.1016/j.nima.2010.06.187

Web


Abstract
Treatment of ocular cancers using eye plaque brachytherapy is now an established medical procedure. However, current QA for these eye plaques is quite rudimentary, limiting the opportunities for precise per-tumour plaque customisation. This paper proposes and experimentally validates a new technique for imaging of eye plaque dose distributions using a high-resolution pixelated silicon detector. Results are presented demonstrating the 2D and 3D isodose surfaces produced using experimental data collected using this method.
Grants

Projects


Cite article as:
M. Weaver, J. Green, M. Petasecca, M. Lerch, D. Cutajar, D. Franklin, J. Jakůbek, M. Carolan, M. Conway, S. Pospíšil, T. Kron, P. Metcalfe, M. Zaider, A. Rosenfeld, "Three-dimensional dosimetry imaging of I-125 plaque for eye cancer treatment", Nucl. Instr. Methods A, Vol. 633, Suppl. 1, p. S276-S278, doi:10.1016/j.nima.2010.06.187 (2011)