Neutron Imaging with MEDIPIX2 Chip and a Coated Sensor
Autor
Rok
2006
Časopis
NIM A Vol. 560, pages 143-147
Web
Obsah
detector device Medipix-2 with a silicon sensor chip (256 x 256 square pixels of 55 m size) can be adapted for neutron detection by deposition of a converter layer onto its surface. Thermal neutrons are converted in the layer into secondary radiation which is detected by the pixel detector. Spatial resolution, detection efficiency and other properties were simulated for different converter materials and verified experimentally. The performance of the imager was compared with several types of neutron imaging systems. The results demonstrate superior properties of the Medipix-2 based neutron imager namely in terms of spatial resolution, linearity and dynamic range. The system is very promising in applications such as neutron micro-tomography
Projekty
Ocenění
Invited contribution to conference VERTEX 2004, Como, Italy
Příklad citace článku:
J. Jakůbek, T. Holý, E. Lehmann, S. Pospíšil, J. Uher, D. Vavřík, "Neutron Imaging with MEDIPIX2 Chip and a Coated Sensor", NIM A Vol. 560, pages 143-147 (2006)