Quality of X-ray transmission radiography based on single photon counting pixel device

Autor

Rok
2005

Časopis
NIM A Vol. 546, pages 113-117

Web


Obsah
This work devotes to the optimization of digital X ray transmission radiography for the observation of time dependent processes using a single X-ray photon counting pixel device. The maximization of frame rate of acquired roentgenograms is crucial for this purpose. The influence of the X ray tube voltage on exposure time, which is necessary for sufficient quality of roentgenograms, was investigated. Roentgenograms are distorted due to non uniform pixel detection efficiency across the X ray detector. Such distortion can be successfully eliminated by the proposed “Direct thickness calibration method” regardless of X ray beam hardening, which is due to the variation of inspected body absorption.
Granty

Projekty


Příklad citace článku:
J. Jakůbek, D. Vavřík, S. Pospíšil, J. Uher, "Quality of X-ray transmission radiography based on single photon counting pixel device", NIM A Vol. 546, pages 113-117 (2005)